Optimising the IAEA remote and automatic quality control for digital radiography: phantom design and reproducibility investigation

Document Type

Article

Publication Date

1-1-2026

Abstract

This study evaluated the reproducibility and uncertainty of locally manufactured International Atomic Energy Agency (IAEA) remote and automated quality control (QC) phantoms for digital radiography (DR). Four investigations assessed (I) copper (Cu) target cutting method, (II) aluminium (Al) target thickness variation, (III) Cu target orientation, and (IV) phantom reproducibility and inter-system performance of identical phantoms fabricated in a single workshop. An uncertainty budget was developed following IAEA Human Health Series No. 39 to quantify variability. Different Cu cutting methods produced comparable modulation-transfer-function (MTF) values, with machining-related uncertainty up to 11%. The 0.1 cm × 4 Al configuration most closely matched the IAEA nominal thickness, measuring only 1% greater than the recommended value. Rotating the Cu target introduced variability in MTF, with coefficients of variation (COV) ranging from 6 to 11%, while reproducibility testing across DR systems showed COV values between 1 and 7% across image quality metrics measured. These findings demonstrate that standardised single-workshop fabrication can produce reproducible and reliable IAEA DR phantoms, supporting the feasibility of the IAEA remote QC programme as a cost-effective and sustainable approach to radiography quality assurance.

Keywords

Digital radiography, Image quality metrics, Quality control phantom, Reproducibility and uncertainty

Publication Title

Physical and Engineering Sciences in Medicine

ISSN

26624729

DOI

10.1007/s13246-026-01708-4

Share

COinS